Abstract

Estimation of low-density lipoprotein cholesterol (LDL-C) using the Friedewald (FR) formula is often inaccurate when triglycerides are elevated or VLDL particle composition is altered. We hypothesized that LDL-C estimation by the FR formula and other measurement methods might also be inaccurate in individuals treated with a cholesteryl ester transfer protein (CETP) inhibitor. An assay comparison study was conducted using pre and posttreatment serum samples from 280 of the 811 patients treated with the CETP inhibitor anacetrapib in the DEFINE study (determining the efficacy and tolerability of CETP inhibition with anacetrapib). After 24 weeks of treatment with anacetrapib, mean LDL-C values by FR formula, Roche direct method (RDM) and Genzyme direct method (GDM) deviated from that measured by the β-quantification (BQ) reference method by -12.2 ± 7.5, -10.2 ± 6.6, -10.8 ± 8.8 mg/dl, respectively. After treatment with anacetrapib, the FR formula and detergent-based direct methods provided lower LDL-C values than those obtained by the BQ reference method. The bias by the FR formula appeared to be due to an overestimation of VLDL-C by the TG/5 component of the formula. Evaluation of the clinical significance of these findings awaits comprehensive lipid and cardiovascular outcome data from ongoing Phase III clinical studies of anacetrapib.

Highlights

  • Estimation of low-density lipoprotein cholesterol (LDL-C) using the Friedewald (FR) formula is often inaccurate when triglycerides are elevated or VLDL particle composition is altered

  • To evaluate LDL-C measurement methods across a wide range of LDL-C values, the patient subset was selected according to baseline plasma LDL-C values, as calculated by the FR formula using the original plasma lipid measurements performed during the course of the study [10]

  • The patient population subset was enriched with a greater percentage of patients with lower baseline LDL-C values

Read more

Summary

Introduction

Estimation of low-density lipoprotein cholesterol (LDL-C) using the Friedewald (FR) formula is often inaccurate when triglycerides are elevated or VLDL particle composition is altered. We hypothesized that LDL-C estimation by the FR formula and other measurement methods might be inaccurate in individuals treated with a cholesteryl ester transfer protein (CETP) inhibitor. After 24 weeks of treatment with anacetrapib, mean LDL-C values by FR formula, Roche direct method (RDM) and Genzyme direct method (GDM) deviated from that measured by the ␤-quantification (BQ) reference method by –12.2 ± 7.5, –10.2 ± 6.6, –10.8 ± 8.8 mg/dl, respectively. The FR formula and detergent-based direct methods provided lower LDL-C values than those obtained by the BQ reference method. Measurement of LDL-C after treatment with the CETP inhibitor anacetrapib.

Objectives
Methods
Results
Conclusion

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.