Abstract
Abstract This paper reported the method development and validation of a gas chromatography with thermal conductivity detector (GC-TCD) method for the measurement of the gaseous products of hydrodeoxygenation (HDO). The method validation parameters include selectivity, precision (repeatability and reproducibility), accuracy, linearity, limit of detection (LoD), limit of quantitation (LoQ), and robustness. The results showed that the developed method was able to separate the target components (H2, CO2, CH4 and CO) from their mixtures without any special sample treatment. The validated method was selective, precise, accurate, and robust. Application of the developed and validated GC-TCD method to the measurement of by-product components of HDO of bio-oil revealed a good performance with relative standard deviation (RSD) less than 1.0% for all target components, implying that the process of method development and validation provides a trustworthy way of obtaining reliable analytical data.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.