Abstract

This paper describes the measurement of microscopic radio frequency (RF) field distribution with a magnetic force microscope (MFM) tip exploring the beat signal from a coplanar waveguide (CPW) with a signal line as fine as 5 μm and a ground line of 50 μm. To produce a beating field in close proximity to the CPW, two RF currents with slightly different frequencies are supplied, where one is the signal current with a fixed frequency of 1.1 GHz and the other is the reference. The reference is adjusted to produce a beat with a frequency near the cantilever mechanical resonance frequency. Thus, the mechanical resonance of the cantilever excited by the beat field includes information about the 1.1 GHz field distribution from the CPW signal current. Detection of a beating field with a MFM tip can provide very high resolutions of the RF field distribution in the proximity of RF circuit component.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.