Abstract

The stiffness tester and torque meter were found to yield nearly the same measurements of bending deformation for orthodontic wires as small as .007 inch diameter, provided the different bending apparatus are calibrated to each other. Although frictional losses in the test apparatus did not appear to be significant, the test span length does significantly affect measured values of Young's modulus and flexural yield strength, in a manner similar to that previously found for larger-diameter wires.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.