Rotatable retarder fixed polarizer (RRFP) Stokes polarimeters, which employ uniformly spaced angles over 180° or 360°, are most commonly used to detect the state of polarization (SOP) of an electromagnetic (EM) wave. The misalignment error of the retarder is one of the major error sources. We suppose that the misalignment errors of the retarder obey a uniform normal distribution and are independent of each other. Then, we derive analytically the covariance matrices of the measurement errors. Based on the covariance matrices derived, we can conclude that 1) the measurement errors are independent of the incident intensity s0, but seriously depend on the Stokes parameters (s1, s2, s3) and the retardance of the retarder δ; 2) for any mean incident SOP, the optimal initial angle and retardance to minimize the measurement error both can be achieved; 3) when N = 5, 10, 12, the initial orienting angle could be used as an added degree of freedom to strengthen the immunity of RRFP Stokes polarimeters to the misalignment error. Finally, a series of simulations are performed to verify these theoretical results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.