Abstract

The measurement accuracy of complex permittivity in the dielectric rod resonator method has been investigated. The measurement error of relative permittivity is 0.05% (standard deviation), which is permissible for the applications of resonators. The dielectric loss was measured by setting conducting plates at both ends of the dielectric resonator. The frequency dependence of the conductivity due to the surface roughness of the plates was investigated using three kinds of standard rod resonators. It is concluded that the measurement error of tan δ can be reduced to 1.5×10-6 (standard deviation) for a dielectric specimen with εr=30.2 at 10 GHz by using the high-Q standard dielectric rod resonators which are placed between conducting plates with surface roughness less than the skin depth at the measurement frequency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.