Abstract

With the advent of technology, multi-energy X-ray imaging is promising technique that can reduce the patient's dose and provide functional imaging. Two-dimensional photon-counting detector to provide multi-energy imaging is under development. In this work, we present a material decomposition method using multi-energy images. To acquire multi-energy images, Monte Carlo simulation was performed. The X-ray spectrum was modeled and ripple effect was considered. Using the dissimilar characteristics in energy-dependent X-ray attenuation of each material, multiple energy X-ray images were decomposed into material depth images. Feedforward neural network was used to fit multi-energy images to material depth images. In order to use the neural network, step wedge phantom images were used for training neuron. Finally, neural network decomposed multi-energy X-ray images into material depth image. To demonstrate the concept of this method, we applied it to simulated images of a 3D head phantom. The results show that neural network method performed effectively material depth reconstruction.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.