This paper proposes a new master device for teleoperated needle insertion-type interventional robotic system. The 5-DOF master device is optimally designed based on the newly defined interventional procedures and the physicians' requirements. It comprises a 2-DOF rotational mechanism for adjustment of needle orientation, a 2-DOF translational mechanism for fine-tuning of needle entry point, and a handle assembly. The handle assembly includes a 1-DOF translational mechanism for needle insertion and buttons for operation mode selection. The passive actuation modules of the rotational mechanism and the active actuation modules of the translational mechanism are controlled appropriately for the selected mode according to the procedure phase. The needle insertion mechanism also warns the user by vibrating the shaft when the needle reaches the dangerous region.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.