Abstract
This report reviews the findings of a massive pseudotumor detected pre-operatively in a 13-year-old revision total hip arthroplasty. The case is unique in that the bearing involved was a 28mm zirconia ceramic head on a polyethylene liner. We propose that the pseudotumor arose from ultrafine titanium particles liberated from the proximal porous coating of the femoral stem. We suspect that the osteolysis produced from polyethylene wear exposed the proximal porous coating and, via a process of mechanical abrasion with the surrounding soft tissues, liberated ultrafine titanium particles. We believe the pseudotumor formed because the patient was pre-sensitized to metal debris based upon a pre-operative lymphocyte T-cell proliferation test (LTT). Based upon this unique case, we feel that pseudotumors more likely form when there is a high rate of ultrafine metal particles generated in a pre-sensitized patient. Finally, we introduce what we believe are the main biologic wear responses in THA. Further research is needed to validate this proposed model.Keywords: pseudotumor, ceramic, polyethylene, osteolysis, THA, bearing wear eesponse, titanium debrisLevel of Evidence: AAOS Therapeutic Study Level IV
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.