Abstract

We compared three current methods (immunoinhibition, "Isomune-CK" immunoprecipitation, and the Tandem-E CKMB II immunoenzymometric assay) for determination of creatine kinase (CK; EC 2.7.3.2) isoenzyme MB in serum. Although results inter-correlated well, the immunoinhibition assay gave higher activity values. Atypical CK forms did not interfere with the immunoprecipitation and immunoenzymometric methods. In acute myocardial infarction the catalytic properties of CK decreased with the enzyme's age, as reflected by a steady increase in activation energy of the catalyzed reaction. In septicemia patients with very low CK and CK-MB catalytic activity, mean CK-MB mass concentration exceeded the upper reference limit, suggesting an increased rate of loss of activity concentration in these patients' sera. Because of the assay's lesser susceptibility to conformational changes at the active site of the enzyme, we suggest that measurement of CK-MB mass concentration is better suited for infarct sizing than measurement of catalytic activity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.