Abstract

A novel fast-cured, boron-loaded, plastic organic scintillator was developed and characterized. A diisopropylnaphthalene-containing scintillator was manufactured in minutes with the addition of two commercially available fast-curing monomers, isobornyl acrylate (IBOA) and 1,6-Hexanediol dimethacrylate (HDDMA). The light yield of the device was measured to be approximately 70% relative to a similarly sized, but several-year-aged, EJ-276 detector. Using a coincidence capture technique, the neutron capture light output was found to be 87.7 ± 4.2 keVee, which was 1.47 ± 0.07 greater than the light output from a 59.5 keV electron. Producing a viable fast-cured and thermal-neutron-sensitive detector enables the capability to pursue cost-effective 3D printed thermal neutron detectors in a timely manner.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.