Abstract
Magnetic resonance elastography (MRE) using mechanical stimulation has demonstrated diagnostic value and clinical promise in breast, liver, and kidney at 1.5 Tesla (T). However, MRE at 1.5T suffers from long imaging times and would benefit from greater signal-to-noise for more robust postprocessing. We present an MRE sequence modified for liver imaging at 3.0T. To avoid artifacts in the phase images, the sequence maintains a short TE by using a second harmonic approach, including stronger motion encoding gradients, shorter radio frequency pulses and an echo-planar readout. Scan time was decreased by a factor of approximately 2 relative to 1.5T by using an EPI readout and a higher density sampling of the phase waveform was used to calculate shear stiffness and viscosity. Localized (small region of interest) and global (whole-liver region of interest) measurements in normal healthy subjects compared very favorably with previously published results at 1.5T. There was no significant difference between global and localized measures.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.