Abstract
Continuous glucose monitors (CGMs) are prone to faults termed pressure-induced sensor attenuations (PISAs), particularly when the user rolls over on the sensor during sleep. PISAs result in false, low blood glucose readings, leading to undesirable pump shutoffs and an increased risk of hyperglycemia. Data from an outpatient trial with PISA glucose readings labeled for 1125 nights was used. Machine learning methods such as decision trees, support vector machines, neural networks, random forest, and gradient-boosted machine models are compared against each other and with a previously reported rules-based algorithm for PISA detection on the same data set. The best-performing gradient-boosted machine model is further improved using the developed start-PISA model. Model interpretation methods are used to confirm that PISA behavior is well-captured and provides insights into decision-making.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.