Abstract

In this paper, a novel method for lumen boundary identification is proposed using Neutrosophic c_means. This method clusters pixels of the intravascular optical coherence tomography image into several clusters using indeterminacy and Neutrosophic theory, which aims to detect the boundaries. Intravascular optical coherence tomography images are cross-sectional and high-resolution images which are taken from the coronary arterial wall. Coronary Artery Disease cause a lot of death each year. The first step for diagnosing this kind of diseases is to detect lumen boundary. Employing this approach, we obtained 0.972, 0.019, 0.076 mm2, 0.32 mm, and 0.985 as mean value for Jaccard measure (JACC), the percentage of area difference (PAD), average distance (AD), Hausdorff distance (HD), and dice index (DI), respectively. Based on our results, this method enjoys high accuracy performance.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.