Abstract

The erbium laser could provide an interesting solid state alternative to the excimer laser (193 nm) for photokeratectomy. The very high absorption of the erbium laser radiation (2.94 micrometers ) by water leads to precise excisions in the cornea with limited thermal damage. In this work we have studied the effects of Q-switch erbium pulses on rabbit eyes. An ultra-fast imaging technique, based on an intensified video camera and on a nitrogen pumped dye laser, was used to investigate the dynamic evolution of the plume during the laser interaction. During the interaction with eyes deprived of the epithelium the ratio between vapor and fragments in the plume strongly depends on the hydration of the stroma. Histologic analysis was carried out to estimate the quality of incision and the amount of thermal damage.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.