Abstract
Smaller, lower-profile plates for metacarpal fixation may have the potential to reduce extensor tendon irritation and adhesions, but their sufficiency for stabilizing metacarpal fractures has not been studied. We investigated the relative stiffness and strength of low-profile and conventional plating systems. For apex dorsal bending (bending closed), no plates broke or had notable plastic deformation. The conventional plates exhibited higher overall bending rigidity than all other plates, but had a lower maximum bending moment than the smaller plates. In apex volar bending (bending open) and torsion, the conventional plates were remarkably more rigid and developed remarkably higher torque. In vivo metacarpal loads are primarily apex dorsal bending, and all plates performed well in this mode. Thus, the smaller, low-profile plates may be sufficient for metacarpal fixation, although patient compliance and the use of supplemental stabilization with a cast or splint should be considered.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.