Abstract

Internationally, respirable crystalline silica (RCS) occupational exposure limits (OELs) are being reassessed and, in some jurisdictions, lowered, putting pressure on the capabilities of the analytical techniques used to achieve robust analyses and reliable detection limits. In preparation of a lower OEL, options for lowering the limit of detection (LoD) for RCS analysis have been assessed. Using a Direct-on-Filter X-Ray Diffraction (XRD) analysis under reduced scan speeds in combination with low-noise RCS sampling filters, an LoD of 0.25 µg/filter and a limit of quantification (LoQ) of 0.82 µg/filter can be achieved. Both limits would translate in an LoD of 0.24 µg/m3 and LoQ of 0.78 µg/m3 when sampling respirable dust for 8h at 2.2L/min, providing a technical solution to monitor exposures at the proposed OEL of 0.025mg/m3 (25 µg /m3) and below, with general sampling conditions as typically applied in Australia. This is the first report showing that the OEL of 0.025mg/m3 (25 µg /m3) is measurable by one of the standardized, direct-on-filter XRD methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.