Electrical impedance tomography (EIT) has many potential applications in medicine. The sensitivity and spatial resolution of EIT can be enhanced significantly by increasing the measurement signal-to-noise ratio (SNR), to which a significant contribution is made by the SNR of the current excitation sub-system. In this paper we present a prototype EIT current excitation sub-system with 80 dB SNR, exploiting both digital and analogue techniques. It uses digital waveform synthesis, a 16-bit DAC and subsequent reconstruction filter, to drive an enhanced Howland current source. Detailed analysis and testing of the current sub-system are presented. Its output impedance is 10 MΩ for different load impedances, varying slowly over the frequency range from 10 kHz up to 4 MHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.