Measurement of in vivo chlorophyll a fluorescence at temperatures lower than 20 degrees C can cause an artifactual, nonphotochemically related overestimation of variable fluorescence leading to the calculation of negative values for the nonphotochemical quenching parameter and an underestimation of the photochemical quenching parameter. This artifact is observed only upon exposure of the leaf sample to actinic light. We suggest that a temperature differential between the fiber-optic probe and the leaf sample results in the deposition of water vapor on the probe that distorts the light path such that an increased modulated fluorescence signal is observed. This artifact is eradicated by ensuring that the end of the fiber-optic probe is kept free of condensation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.