Abstract

As intraoperative use of electronic equipment becomes more common, familiarity with the potential for electrocardiographic artifact generated by these instruments is increasingly important. The authors' patient underwent endoscopic dacryocystorhinostomy under general anesthesia for treatment of acute on chronic dacryocystitis secondary to acquired nasolacrimal duct obstruction. She had previously had pacemaker placement for treatment of sick sinus syndrome and low ejection fraction heart failure, and use of a microdebrider intraoperatively resulted in loss of pacer spikes and interference signal on electrocardiographic leading to placement of a magnet. Post-operatively, examination of the microdebrider cord revealed a tear that likely lead to the artifactual electrocardiographic signal influencing the decision for magnet placement. Although there were no long-term ill effects, placement of the magnet was an unnecessary intervention that may have been avoided by prompt recognition of artifactual changes that may be induced by intraoperative use of electronic instrumentation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.