Abstract
The intensities of the γ-rays and K X-rays associated with the decay of 249Cf, 243Am, 239Np and 243Cm were measured with respect to IAEA standards. Because of their long half-lives and ease of standardization by accurate alpha counting, these nuclides are recommended as standards for absolute efficiency calibration of Ge(Li) detectors.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.