Aseptic loosening of cemented knee arthroplasties has encouraged development of uncemented fixation. The Miller-Galante I (MG I) prosthesis was designed to achieve permanent stability through ingrowth into a titanium fiber mesh. Thirty-five knees in 30 patients with MG I knee replacements have been followed clinically and radiologically with a mean follow-up of 12 years. Twenty knees were revised with a mean follow-up of 5 years. Patellofemoral problems, especially avulsion of the polyethylene from the metal-backed patella and in some cases severe metallosis, have been the main reason for revision. Metal-backed patellar component should be avoided. In addition, instability has been a problem. However, the fixation of the components has been excellent, with a high degree of osseous ingrowth displayed at histological analysis of retrieved components. The clinical and radiological results after revision are in most cases good, despite the femoral component having been left in situ.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.