Abstract

A bioanalytical assay for the mutated BRAF inhibitor vemurafenib was developed and validated. For the quantitative assay, human plasma samples were pre-treated using protein precipitation with water-acetonitrile (1/3, v/v) containing sorafenib as internal standard. The extract was directly injected into the chromatographic system. This system consisted of a sub-2μm particle, trifunctional bonded octadecyl silica column with isocratic elution using 0.01% (v/v) of formic acid in a mixture of water and methanol. The eluate was transferred into the electrospray interface with positive ionization and the analyte was detected in the selected reaction monitoring mode of a triple quadrupole mass spectrometer. The assay was validated in a 0.1–100μg/ml calibration range. Within day precisions were 1.6–3.2%, between day precisions 2.7% and 8.2% and accuracies were between 99% and 106% for the whole calibration range. The drug was stable under all relevant conditions. Finally, the assay was successfully used to assess drug levels in a pharmacokinetic mouse study.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.