A rapid, sensitive, and specific liquid chromatography‐electrospray ionization mass spectrometric (LC‐ESI‐MS) method has been developed for quantification of gliclazide in human plasma. The analyte and tolbutamide (internal standard, I.S.) were extracted from plasma samples with n‐hexane–dichloromethane (1:1, v/v) and analyzed on a C18 column. The chromatographic separation was achieved within 4.0 min by using methanol–0.5% formic acid (80:20, v/v) as mobile phase and the flow rate was 1.0 mL/min. Ion signals m/z 324.0 and 271.0 for gliclazide and internal standard were measured in the positive mode, respectively. The method was linear within the range of 2.5–2000 ng/mL. The lower limit of quantification (LLOQ) was 2.5 ng/mL. The intra‐ and inter‐day precisions were lower than 2.8% in terms of relative standard deviation (RSD). The inter‐day relative error (RE) as determined from quality control samples (QCs) ranged from −1.93% to 1.85%. This validated method was successfully applied for the evaluation of pharmacokinetic profiles of gliclazide modified‐release tablets in 20 healthy volunteers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.