Abstract

A liquid chromatographic tandem mass spectrometric method for the determination of trandolapril in human plasma has been developed and fully validated. The article describes, in detail, the bioanalytical procedure and summarizes the validation results obtained. The samples were extracted using HLB Oasis solid-phase extraction cartridges. The chromatographic separation was performed on X-Terra C8 MS column (150 mm × 4.6 mm i.d., 5 μm particle size) using a mobile phase consisting of acetic acid 20 mM and triethylamine 4.3 mM/acetonitrile (40:60 (v/v)), pumped isocratically at 0.35 ml/min. The analytes were detected using a micromass quattro micro triple quadrupole mass spectrometer with positive electrospray ionization in multiple reaction-monitoring (MRM) mode. Tandem mass spectrometric detection enabled the quantitation of trandolapril down to 2.0 ng/ml. Calibration graphs were linear ( r better than 0.996, n = 9) in the concentration ranges 2.0–750 ng/ml and the intra- and inter-day R.S.D. values were less than 3.83 and 3.86% for trandolapril.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.