Abstract

A liquid chromatographic (LC) method is described for the determination of vitamin D3 in sheep plasma. Samples are extracted by one of 2 different methods, depending on the concentration of vitamin D3. The samples are purified by using either a Sep-Pak silica cartridge or a small alumina column, followed by additional cleanup on a Metalsorb LC column. Final analysis was carried out on a 5 micron C18 column using a radial compression separation system with an acetonitrile-methanol solvent system. Vitamin D3 was completely resolved from any interfering compounds in the plasma; total run time was less than 15 min, using a variable wavelength detector set at 264 nm. The method was successfully applied to samples at levels of 1-10 ng added vitamin D3 mL sheep plasma, with recoveries in the range 90-97%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.