Abstract
We report a simple, efficient method for removing dexamethasone from plasma by use of solid-phase extraction columns. The dexamethasone is then quantified by "high-performance" liquid chromatography with ultraviolet detection at 254 nm. Dexamethasone concentrations and peak-height ratios were linearly related over the range 5 to 200 micrograms/L. The lower limit of sensitivity was 5 micrograms/L. The average recoveries of dexamethasone and methylprednisolone (internal standard) were 72 and 71%, respectively. This procedure offers improved efficiency over that of previously described methods by decreasing analytical time and improving sample clean-up. Sensitive, specific, and relatively inexpensive, this method is suitable for most clinical applications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.