Abstract

Reconstructing the integral of a distribution of radioactivity along a line from Compton camera data could be used to produce a parallel projection, or to perform tomosynthesis, or to reconstruct the whole three-dimensional distribution itself. Analytic methods for reconstructing an integral of radioactivity along a line are presented here. The sets of data that allow an integral along a line and a parallel projection of the distribution to be reconstructed are determined here by interpreting these methods from a geometric viewpoint. These methods and the sets of data depend upon which of the two models is assumed for the data. Both of these models have been previously proposed by other researchers. In addition, a new camera design is proposed here that makes it possible to measure all these sets of data. In this design, a first detector element has to be seen from the second detector in only a "semicircle of direction." Also, two techniques for increasing the sensitivity of the new camera design are proposed here. Computer simulations are performed to illustrate these reconstruction methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.