Abstract

Distraction osteogenesis is a powerful tool for addressing segmental defects and limb-length discrepancies. Guidelines for the rate and rhythm of distraction have been described. The possibilities of early consolidation or nonunion threaten clinical success. A quantifiable method for monitoring the distraction gap would be useful. Previous methods to measure regenerate callus stiffness have not gained widespread clinical acceptance, largely because of cumbersome instrumentation. A rabbit tibial lengthening model was used to show the clinical utility of a digital torque wrench in monitoring axial stiffness of the regenerate bone callus during distraction osteogenesis. We confirmed the linear increase in peak torsional stiffness with time, which has been reported by others. This relationship may prove to be a useful clinical adjunct in guiding the rate and rhythm of distraction during limb lengthening.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.