Abstract

This paper describes length measurements using a regular crystalline lattice as a reference scale and a dual tunnelling unit scanning tunnelling microscope (DTU-STM) as a detector in a thermo-stabilized cell. Direct length comparisons between a certified standard scanning electron microscope (SEM) grating with an average pitch of 240 nm and a highly oriented pyrolytic graphite (HOPG) lattice spacing, which is 0.246 nm, were performed. Images of the grating and the HOPG were simultaneously obtained in the range of m using the DTU-STM. The thermo-stabilized cell was developed to suppress any thermal drift error. In order to shorten the measurement time and thus reduce the thermal drift error, lengths of m for the two samples were measured along the fast scanning axis. Rapid scanning, in which the tip speed was up to , was also utilized to shorten the measurement time. The grating pitch obtained by comparison with the HOPG lattice spacing agreed with that calibrated by the conventional diffraction method within an error of 2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.