Abstract

Distinct architectures of layer-by-layer (LbL) films made of carbon nanotubes and gold nanoparticles were investigated to serve as the matrix to immobilize cholesterol oxidase, with which cholesterol could be detected using amperometry. The gold nanoparticles were synthesized by metal reduction stabilized in poly(allylamine hydrochloride) (PAH) providing a stable AuNPs-PAH suspension, while multi-walled carbon nanotubes (CNTs) were functionalized with carboxylic groups to obtain an aqueous suspension. The LbL films were deposited on ITO, with a cushion film of PAH and poly(vinyl sulfonic acid) (PVS). Owing to the synergy between CNTs and AuNPs, the electrode ITO/(PAH/PVS)₂ (AuNPs-PAH/CNTs)10 was selected for immobilization of cholesterol oxidase (ChOx). This sensor could detect cholesterol with a limit of detection of 14.8 μmol L-1 and sensitivity of 36.47 (μA cm-2)·(mmol L-1)-1. It was also able to determine cholesterol in egg yolk with a recovery of 97.7%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.