Abstract

There are several noninvasive techniques for assessing the kinetics of tear film, but no comparative studies have been conducted to evaluate their efficacies. Our aim is to test and compare techniques based on high-speed videokeratoscopy (HSV), dynamic wavefront sensing (DWS), and lateral shearing interferometry (LSI). Algorithms are developed to estimate the tear film build-up time T(BLD), and the average tear film surface quality in the stable phase of the interblink interval TFSQ(Av). Moderate but significant correlations are found between T(BLD) measured with LSI and DWS based on vertical coma (Pearson's r(2)=0.34, p<0.01) and higher order rms (r(2)=0.31, p<0.01), as well as between TFSQ(Av) measured with LSI and HSV (r(2)=0.35, p<0.01), and between LSI and DWS based on the rms fit error (r(2)=0.40, p<0.01). No significant correlation is found between HSV and DWS. All three techniques estimate tear film build-up time to be below 2.5 sec, and they achieve a remarkably close median value of 0.7 sec. HSV appears to be the most precise method for measuring tear film surface quality. LSI appears to be the most sensitive method for analyzing tear film build-up.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.