Abstract

Laser ablation inductively coupled plasma mass spectrometry (LA-ICP-MS) was used as a powerful multielement analytical method for trace analysis of geological glasses which are useful as reference materials for geochemical in-situ microanalytical work. The quantification of the analytical results was carried out using the BCR-2G and NIST 612 glass standard reference material (SRM). The experimentally determined relative sensitivity coefficients (RSC) for both SRMs vary between 0.2 and 3 for most of the elements, with increasing mass an increasing of relative sensitivity coefficients was observed. The relative standard deviation (RSD) for determination of trace element concentration of most elements (N=3) are between 2 and 10%. The determination of trace elements in various geological glasses by LA-ICP-MS yielded a good agreement with the reference values and those results of other trace analytical methods.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.