Abstract

While benefits of neurosurgical intraoperative ultrasound (IOUS) are reported frequently, this method still has some significant pitfalls, which are described less often. However, sufficient knowledge on dealing with IOUS drawbacks, particularly various image artifacts, is important for successful surgery. We report a case of failed IOUS-guided pediatric cerebellar pilocytic astrocytoma resection, incorrectly evaluated as gross total resection according to IOUS. A large tumor residuum was left in place. Successful IOUS-guided reoperation using new IOUS technology and appropriate ultrasound imaging technique are described. The most probable reasons for initial resection failure and crucial points of reoperation, predominantly dealing with IOUS artifacts, are discussed. Neurosurgeons should be aware of IOUS limitations and have sufficient knowledge about how to overcome them before adopting routine use of this intraoperative imaging modality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.