Abstract

The design and construction of a cylindrical proportional counter 36 in. long, 12 in. i.d. is described, and its performance as a spectrometer for studying soft radiations from radioactive samples of low specific activity is given. The counter tank is operated in an axial magnetic field of 2500 gauss to constrain beta rays to helical paths. A line width of 3.4% (ΔE/E) is obtained for 625-kev electrons from a Cs137 source mounted on a thin film placed within the tank. Simple theory would predict 2.2%. Beta spectra are readily obtained with sources having total activities less than 10 mμC.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.