Abstract

Current standard EN 50463-2 indicates the tests and the requirements to be satisfied for an energy measurement system of a traction unit for railway applications. Some of these tests are to be done with several harmonics superposed on the rated voltage, respectively current. However, no calibration systems satisfying the standard requirements were available few years ago. The work performed in the EURAMET project “MyRailS” leads to the development of fictive power sources and reference measurement systems described in this paper. Therefore, it is possible to generate distorted 25 kV-50 Hz voltages with harmonics up to 5 kHz and 90° phase-fired currents up to 500 A with harmonics up to 5 kHz. The generated power is measured by developed traceable reference systems with accuracy better than 0.5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.