Abstract

We have studied the biomechanical stability in vitro of three different Kirschner (K) wire configurations in three types of simulated scaphoid waist fractures. The fractures were created with a saw in Sawbones models. There were three fracture patterns: perpendicular to the long axis of the scaphoid model; and 30° and 20° oblique to that. Two 1.2 mm. K-wires were used in each scaphoid. The three configurations were: parallel; 20° oblique; and crossing. The oblique or crossing configurations of K-wires were the most stable depending on the fracture pattern.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.