Abstract

Percutaneous needle-based procedures such as prostate brachytherapy demands for accurate placement of the needle tip at target locations. Recently, robotic needle insertion systems have been made available to help physicians in needle guidance and control inside tissue. It is often challenging to obtain an accurate and real-time position of the needle tip in clinical practice using medical imaging techniques. However, this information is vital for closed-loop control of the needles inside tissue. This work presents an SMA-actuated active flexible needle that is controlled inside a phantom without a need for a position sensor or a medical imaging device. The needle tip position feedback is found using shape sensing capabilities of the embedded SMA-wire actuators and a force sensor at the needle base. Three models were characterized and used to estimate needle tip position in real time. The control scheme was then tested on the active flexible needle to track a desired triangular trajectory in a phantom. It was shown that the control scheme presented in this work was able to manipulate the needle in this path with a reasonable accuracy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.