Abstract

We previously demonstrated that peripheral blood (PB) is a reliable source for testing JAK2V617F mutation in patients with myelofibrosis (MF); saliva has also been tested to detect such mutation, however its diagnostic accuracy as compared to PB has not been validated.In this study, we prospectively tested 167 patients with MF for JAK2V617F mutation, using both saliva and PB collected at the same time from each patient. The concordance between the 2 sources was 96%, with a sensitivity of 100% and a specificity of 90%. The only factor associated with false positivity on saliva was ongoing transfusion dependency.JAK2V617F testing using saliva is a simple, non-invasive, and potentially a more reliable method than PB for measuring JAK2 status and assessing V617F allelic burden in patients with transfusion dependency.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.