Abstract

A new candidate reference method is presented for the determination of thyroxine in serum. The method is based on isotope dilution-liquid chromatography/tandem mass spectrometry using electrospray for ionization. The internal standard used was 13C6-thyroxine, sample pretreatment consisted of protein precipitation and a two-step liquid/liquid extraction procedure, HPLC was performed on a C-18 column with an eluent containing methanol/water/formic acid (60:40:0.1, by volume), and finally thyroxine and its isotopically labeled analogue were measured in the selected reaction monitoring mode for the transitions m/z 777.7--> 731.7 and m/z 783.7--> 737.7, respectively. The detection limit for thyroxine was 6 pg, the within-run coefficient of variation was 1.1%. The samples were measured in six-fold: in duplicate on three independent days. The mean overall coefficient of variation of the method was 1.6%. The new method was evaluated by measuring nine control sera previously determined by an existing ID-GC/MS method. The differences between the results of the two methods ranged from-1.6% to +3.3%, with a mean of +0.2%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.