Abstract

This study aimed to test the validity of an automatic oscillometric device to measure the blood pressure (BP) in children (n = 191) and adolescents (n = 127) aged 3 to 18 years. Systolic BP (SBP) and diastolic BP (DBP) levels were measured simultaneous by automatic device and mercury column with Y-connection. To verify the validity, Bland-Altman plots and limits of agreement of 95% (95% LOA), specificity and sensitivity of the device, and the grade of British Hypertension Society (BHS) criteria were used. The monitor measurements demonstrated lower measurement bias (mean difference [95% LOA]): 1.4 (-9.9 to 12.8) mmHg in children and 4.3 (-7.8 to 16.5) mmHg in adolescents for SBP. For DBP, it was 2.2 (-7.4 to 11.7) mmHg in children and 1.4 (-8.4 to 11.1) mmHg in adolescents. The sensitivity in children was 21.4 (95% CI = 16.3-26.6), and in adolescents, it was 20.0 (95% CI = 13.2-26.8); the specificity was 95.9 (95% CI = 93.4-98.4) in children and 100.0 (95% CI = 100.0-100.0) in adolescents. The monitor-tested ratings are Grade B for SBP in children and SBP and DBP in adolescents and Grade C for DBP in children. The automatic monitor presented high values of specificity and lower values of sensitivity to the diagnosis of HBP; however, it can be considered accurate (lower measurement bias) and valid for epidemiological and clinical practice in accordance with BHS criteria.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.