Abstract
Optimization of the Varian Saturn 2200 ion-trap tandem mass spectrometry (IT-MS/MS) system and comparison of its data quality with two other detection methods [electron-capture detection (ECD) and high-resolution mass spectrometry (HRMS)] was pursued by measuring polychlorinated biphenyls (PCBs) levels in fish and shellfish samples. IT-MS/MS methodology provided limits of detection (LOD) comparable to those obtained by ECD but superior specificity for the detection of a selected number of 39 PCB native congeners and 9 13C-labelled PCB standards. The method detection limits (MDLs) established for IT-MS/MS ranged between 1.0 and 5.0 pg/g on a wet weight basis while those obtained by ECD and HRMS were 1.0–4.0 pg/g and 0.1–2.0 pg/g, respectively. Overall, the results obtained in the study demonstrate that gas chromatography (GC) combined with IT-MS/MS provide higher data quality than those achievable by GC–ECD. For this particular set of target analytes the specificity achievable with IT-MS/MS was comparable to that obtained by HRMS and both techniques provided comparable data in terms of accuracy and precision.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.