The ion-optical design of a multiple reflection time-of-flight mass spectrometer and isobar separator (MR-TOF-MS) is presented. The MR-TOF-MS has been developed for the research with exotic nuclides at accelerator facilities and has been designed to provide a mass resolving power (FWHM) exceeding 300,000 and a mass accuracy of down to 10−7. High-performance axially symmetric four-electrode gridless ion mirrors have been developed and optimized for the device. In addition, the MR-TOF-MS comprises several further novel ion-optical elements such as an isochronous post-analyzer reflector and an energy bunching and deceleration system for ion injection into an accumulation trap. The device has been built and commissioned and shows superb performance characteristics in accordance with the design criteria.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.