Abstract
The feasibility of coupling an electron-cyclotron-resonance (ECR) plasma-processing reactor directly to an omegatron mass spectrometer is demonstrated. The ECR plasma is created in a chamber that is coupled to the omegatron through a small, grounded orifice. Ions created in the ECR chamber flow along the magnetic field into the omegatron analysis cell, and the mass spectrum of these ions is recovered. Using this technique the mass spectra of both single-component (He) and two-component (N/sub 2/ and N) ECR plasmas were measured. The mass resolution as a function of the omegatron excitation voltage for He and N/sub 2/ plasmas was obtained and found to compare well to theoretical calculations.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.