Abstract

High density concrete is usually used for radiation shielding around radiotherapy treatment rooms. Because the concrete is specified differently at the design, construction, and verification stages, the relationship between the intended performance and the actual performance of the shielding material might not be entirely clear. In this study, cylindrical samples of high density shielding concrete were taken as each section of a new radiotherapy bunker was poured. The shielding performance of each sample [measured by beam attenuation and tenth-value layers (TVL)] was evaluated for 15 MV and 6 MV x-ray beams and for the 1.25 MeV monoenergetic gamma beam from a Co source. Transmission curves to 3 TVL were mapped for a representative sample. The samples were also imaged and analyzed using Co Cone Beam Computed Tomography (CoCBCT). Results indicate no significant change in the TVL of high density concrete samples as they cure. The minor fluctuations in shielding properties observed are explained by the heterogeneous structure of the samples as indicated in the CoCBCT images.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.