Abstract

The article gives the results of research of electrospray stability with the use of a dynamic splitting of a flow in the spray area under normal conditions. An acidified water-acetonitrile solution (1: 1) was used as an eluant. Experimental results for the used configuration of electrospray junction with a coaxial selection of fluid excess as a steam-gas mixture taken out of the dispersion area (meniscus fluid on the butt-end of a capillary) in the compass of a fluid flow along the capillary from 50 up to 200 μL/min showed more stable relation of an atomized solution flow in comparison with a mode without use of a dynamic splitting of an atomized fluid flow.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.