Abstract

This study investigates the contrast behavior of 2D inversion recovery (IR) prepared turbo spin echo (TSE) sequences that use RF pulse schemes with variable low flip angles (hyperTSE) to reduce RF power deposition. A framework of equations and calculations for adapting the sequence parameters is presented by which equivalent image contrast is achieved compared to conventional IR-TSE imaging. Although the inversion time (TI) and repetition time (TR) do not need to be changed, the echo time (TE) has to be prolonged such that the effective TE (TE(eff)) is preserved. Measurements in healthy volunteers confirmed this finding for IR-TSE sequences using different TIs: fluid attenuated inversion recovery (FLAIR), gray matter (GM)-white matter (WM)-IR, and short tau IR (STIR). The results demonstrate that hyperTSE sequences enable high-quality IR-prepared imaging with a considerably reduced specific absorption rate (SAR).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.