Abstract

It is important to measure organ dose in diagnostic radiology & radiotherapy. Because measurement on patients has some limitations, phantoms are being constructed to be implemented. One kind of several phantoms which is being used so far is anthropomorphic phantom which is similar to a man. In this work, using human natural bone as phantom skeletal tissue & paraffin plus NaCl (as impurity) as soft tissue substitute an anthropomorphic phantom was constructed which consists of three parts: 1. head & neck 2.torso 3.hip. Phantom dimensions were selected from the standard man. To measure absorbed dose in tissues, at the sites of parotid, thyroid, sternum, diaphragm, abdomen and hip some cylinders were inserted which were constructed from phantom material and had some cavities to insert TLDs in different depths. Due to importance of photoelectric effect in diagnostic radiology & Compton effect in radiotherapy, two important parameters which should be considered in construction of a phantom is effective atomic number and electronic density were6.57& 3.36×10 23 electron.gr -1 , respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.