Abstract
This study describes the use of discrete, highly selective covering inner membranes in amperometric enzyme electrodes for direct biofluid analysis. Such inner membranes when interposed between the enzyme layer and the electrode working surface has been found to be both selective, as well as biocompatible. Cellulose acetate or detergent (Triton X-100 and Tween 80) modified cellulose acetate and polyethersulphone polymeric membranes, were both found to be highly selective against electroactive solutes normally present in biofluids. The inclusion of such inner perm-selective membranes greatly enhanced the biocompatibility of the enzyme electrode and reduced the passivation of the working electrode on exposure to whole blood or serum. The study indicates that electrode drift in biological solutions may also be due to passivation of the working electrode by small diffusible surface active species.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Scandinavian Journal of Clinical and Laboratory Investigation
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.