Abstract

We describe a previously unreported phenomenon of intermittent outflow right ventricular tract capture from the atrial lead of a dual-chamber pacemaker. This was more obvious at slower paced atrial rates and disappeared by decreasing the atrial pulses voltage. Electroanatomical mapping showed that the onset of activation was nearly simultaneous at the insertion site of the atrial lead and at an intermediate level of the right ventricular outflow tract. This exceptional finding might be erroneously diagnosed as due to pseudo-pseudo fusion beats.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.